# AGRINOVO-DEN-100: Tuning Fork Density / Concentration Meter (Insertion Type)

Agrinovo technical datasheet, revision 1.2 (October 2026).
PDF version: https://agrinovo.io/datasheets/sensors/AGRINOVO-DEN-100_Datasheet.pdf

## 1. Overview

The AGRINOVO-DEN-100 is a rugged, fully-integrated insertion-type density and concentration meter. An acoustic excitation source drives a metal tuning fork at its free-oscillation centre frequency. The frequency shifts in proportion to the density of the liquid in contact with the fork; temperature drift is compensated by an internal PT100, and concentration is derived from the density relationship at 20°C. The sensor is a "plug-and-measure, maintenance-free" instrument for continuous inline or in-tank measurement.

### Key Features

- Density range: 0 to 1.8 g/cc (0 to 1800 kg/m³)
- Accuracy: ±0.003 g/cc (±3.0 kg/m³)
- Repeatability: ±0.001 g/cc
- Integrated PT100 temperature compensation
- No moving parts  -  low maintenance
- 4 to 20 mA or Modbus RS485 output
- Insertion length up to 2 m
- Wetted materials: 316L / Hastelloy C / Zirconium
- Ex d II BT6 Gb hazardous-area rating
- Suitable for open tanks, closed vessels and pipelines

### Applications

- Concentration control of acids, alkalis and process solutions
- Brine and salinity monitoring
- Fermentation and brewing concentration tracking
- Mass-flow computation (combined with volumetric flow metering)
- Evaporator concentration control and batch reaction end-point detection
- Interface detection in multi-product pipelines

## 2. Density Performance

| Parameter | Specification |
|:----------|:--------------|
| **Range** | 0 to 1.8 g/cc (0 to 1800 kg/m³) |
| **Accuracy** | ±0.003 g/cc (±3.0 kg/m³) |
| **Repeatability** | ±0.001 g/cc (±1 kg/m³) |
| **Temperature influence (compensated)** | ±0.0005 g/cc / °C (±0.5 kg/m³) |
| **Pressure influence (compensated)** | Negligible |
| **Calibration range** | 0.6 to 1.5 g/cc (600 to 1500 kg/m³) |
| **Maximum process viscosity** | 2000 cP |

Temperature-influence figures are the maximum measurement offset caused by process fluid temperature deviating from the factory calibration point. Pressure-influence figures reflect sensitivity and flow response when the process pressure deviates from the calibration pressure.

## 3. Temperature & Pressure

| Parameter | Specification |
|:----------|:--------------|
| **Process temperature** | −20 to +100 °C |
| **Ambient temperature** | −20 to +85 °C |
| **Temperature coefficient (compensated)** | 0.1 kg/m³ / °C |
| **Internal temperature sensor** | PT100 |
| **Maximum working pressure** | 2.5 MPa |
| **Test pressure** | 1.5 × maximum working pressure |

Actual maximum working pressure is limited by the selected process-connection rating.

## 4. Electrical Characteristics

| Parameter | Specification |
|:----------|:--------------|
| **Supply voltage** | 12 to 24 VDC |
| **Supply current** | 100 mA |
| **Output** | 4 to 20 mA (4-wire) or Modbus RS485 |
| **Electrical entry** | M20 × 1.5 |

## 5. Materials & Construction

| Parameter | Specification |
|:----------|:--------------|
| **Wetted parts** | 316L stainless steel, Hastelloy C, Zirconium (selectable) |
| **Tuning-fork finish** | Standard polish or electro-polish |
| **Enclosure** | Aluminium alloy, IP65 |
| **Weight (short rod, stainless)** | 3.5 kg |
| **Weight (long rod)** | Dependent on rod length |

## 6. Process Media Limits

| Parameter | Limit |
|:----------|:------|
| **Particles < 10 μm** | ≤ 40% suspended solids |
| **Particles 10 to 50 μm** | ≤ 20% suspended solids |
| **Entrained gas** | Not generally permitted |
| **Viscosity** | 0 to 2000 cP |

## 7. Hazardous Area Classification

| Certification | Rating |
|:--------------|:-------|
| **Explosion-proof** | Ex d II BT6 Gb |

## 8. Process Connections

The AGRINOVO-DEN-100 is supplied with one of the following process connection options:

- **Flange:** 2" 150LB, 2" 300LB, or DN50 PN25
- **Tri-clamp:** 2" or 3" sanitary clamp
- **Threaded:** G1 male thread

### Dimensional Drawings

**Flange Connection**

**Tri-clamp Connection**

**Threaded Connection**

Standard insertion length is 127 mm. Custom insertion lengths of 100, 200, 300 and 500 mm are available, up to a maximum of 2 m. Housing height is 130 mm, fork rod diameter Ø25 mm.

## 9. Material Compatibility

Guidance for selecting wetted materials against common process media. Use this table as a starting point; confirm against the process conditions of the installation.

| Media | Formula | Concentration | Zirconium | Hastelloy C | 316 SS |
|:------|:--------|:--------------|:---------:|:-----------:|:------:|
| Hydrochloric acid | HCl | 0 to 40% | ★ | ○ | ✕ |
| Sulphuric acid | H₂SO₄ | 0 to 50% | ★ | ○ | ○ |
| Sulphuric acid | H₂SO₄ | 50 to 75% | ○ | ○ | ✕ |
| Sulphuric acid | H₂SO₄ | 75 to 98% | ○ | ○ | ○ |
| Nitric acid | HNO₃ | 0 to 100% | ★ | ○ | ○ |
| Phosphoric acid | H₃PO₄ | 0 to 98% | ✕ | ★ | ○ |
| Sodium hydroxide | NaOH | 0 to 100% | ✕ | ★ | ○ |
| Potassium hydroxide | KOH | 0 to 50% | ★ | ★ | ○ |
| Calcium hydroxide | Ca(OH)₂ | 0 to 50% | ○ | ★ | ○ |
| Urea | (NH₂)₂CO | 0 to 100% | ★ | ★ | ✕ |
| Sodium hypochlorite | NaOCl | 0 to 16% | ✕ | ○ | ✕ |
| Hydrogen peroxide | H₂O₂ | 0 to 90% | ✕ | ★ | ★ |

Legend: **★** recommended &nbsp;·&nbsp; **○** usable within specified concentration and temperature limits &nbsp;·&nbsp; **✕** not suitable

## 10. Installation Notes

**Placement**

- Install where process temperature is stable and free of shock / vibration
- Keep the fork opening perpendicular to flow so sediment and gas cannot build up on the fork
- Fork must remain fully submerged below the minimum liquid level
- Horizontal pipe mounting: open the flange on the pipe **side**, not on top (top openings trap gas)
- Vertical pipe: install on an upward-flowing section
- For pipe flow, keep fluid velocity 0.3 to 0.5 m/s across the fork; if main flow exceeds 1 m/s, use an expansion section or bypass line
- Minimum main-pipe bore: ≥100 mm horizontal / ≥150 mm vertical

**Maintenance & Handling**

- The medium must not freeze on the fork  -  ice damage will destroy the sensing element
- Do not allow sediment to accumulate inside the vessel at the fork location
- Handle gently during installation; do not drop
- Do not operate above rated working pressure
- Do not exceed the specified test pressure when pressure testing
- Do not weld on pipework while the meter is installed
- For slurries or scaling media, select a protective guard, T-sleeve or bell-mouth accessory
- Confirm the selected model is rated for the hazardous-area classification of the installation

## 11. Ordering Information

Build the full part number by concatenating one option from each column.

| Code | Option |
|:-----|:-------|
| **Range** | 1 = 0 to 1.5 g/cm³ · 2 = 0 to 1.8 g/cm³ · X = Other |
| **Process connection** | D1 = 2" 150LB flange · D2 = 2" 300LB flange · D3 = DN50 PN25 flange · K1 = 2" tri-clamp · K2 = 3" tri-clamp · M = G1 thread · XX = Other |
| **Wetted material** | J1 = 316 SS · J2 = Titanium · J3 = Zirconium · J4 = Hastelloy C · JX = Other |
| **Surface finish** | A = Standard polish · B = Electro-polish |
| **Insertion length** | 0 = Standard 127 mm · 1 = 100 mm · 2 = 200 mm · 3 = 300 mm · 5 = 500 mm · X = Other (up to 2 m) |
| **Accessories** | F0 = None · F1 = Mounting bracket · F2 = Bypass pipe · F3 = Flow chamber · F4 = Flush port · F5 = Flange support · F6 = Protective guard · FX = Other |

Accessories are supplied in 316 stainless steel. Accessory dimensions depend on the selected meter configuration.

## Revision History

| Rev | Date | Changes |
|:----|:-----|:--------|
| 1.0 | June 2026 | Initial release |
| 1.1 | August 2026 | Layout revision |
| 1.2 | October 2026 | New layout |

## Contact

Agrinovo, https://agrinovo.io/products, contact@agrinovo.io, +972 54 688 8148 (direct / WhatsApp).
Specifications subject to change without notice.