# AGRINOVO-TUR-110: Self-cleaning Turbidity Sensor

Agrinovo technical datasheet, revision 1.2 (October 2026).
PDF version: https://agrinovo.io/datasheets/sensors/AGRINOVO-TUR-110_Datasheet.pdf

## 1. Overview

The AGRINOVO-TUR-110 is a self-cleaning digital turbidity sensor based on infrared scattered light technology. The infrared light emitted by the source scatters when passing through the sample, and the scattered light intensity is proportional to the turbidity level. A detector positioned at 90° receives the scattered light, and the sensor calculates the turbidity value by analyzing the signal strength.

### Key Features

- Measurement range: 0.01 to 3000 NTU
- Resolution: 0.01 NTU / 0.1 NTU
- Accuracy: ±5% or ±0.5 NTU (whichever is greater)
- Repeatability: ±3%
- Self-cleaning wiper (optional)
- RS485 Modbus-RTU output
- IP68 / NEMA6P protection
- SUS316L stainless steel body

### Applications

- Drinking water treatment inlet/outlet monitoring
- Sedimentation basin turbidity monitoring
- Wastewater treatment process control
- Industrial process water quality monitoring
- Environmental surface water monitoring

## 2. Specifications

| Parameter | Specification |
|:----------|:--------------|
| **Range** | 0.01 to 3000 NTU (customizable) |
| **Accuracy** | ±5% of reading or ±0.5 NTU, whichever is greater |
| **Repeatability** | ±3% |
| **Resolution** | 0.01 NTU or 0.1 NTU (depending on range) |
| **Pressure Rating** | ≤0.2 MPa |
| **Calibration** | Standard solution / sample calibration |
| **Materials** | Body: SUS316L; End caps: PPS + glass fiber; Cable: PUR |
| **Supply Voltage** | 12-24 VDC |
| **Protocol** | Modbus RS485 |
| **Storage Temp** | −15 to 50°C |
| **Operating Temp** | 0 to 45°C (non-freezing) |
| **Dimensions** | Without wiper: 49mm dia. × 222.5mm |
|  | With wiper: 49mm dia. × 230.5mm |
| **Weight** | 0.8 kg |
| **Protection** | IP68 / NEMA6P |
| **Cable Length** | 10 m standard, extendable to 100 m |

## 3. Wiring

| Wire Color | Function | Description |
|:-----------|:---------|:------------|
| **Brown** | V+ | Power Supply (+12V DC) |
| **Black** | AGND | Analog Ground |
| **Blue** | RS485-A | Data+ |
| **White** | RS485-B | Data− |

## 4. Communication Settings

| Parameter | Value |
|:----------|:------|
| **Protocol** | Modbus-RTU |
| **Baud Rate** | 9600 bps (configurable: 4800/9600/19200/38400) |
| **Data Bits** | 8 |
| **Parity** | None |
| **Stop Bits** | 1 |
| **Default Address** | 0x01 (range 1 to 200) |

## 5. Register Map

### Measurement Registers (Function 0x03)

| Address | Description | Type | Length | R/W | Notes |
|:--------|:------------|:-----|:-------|:----|:------|
| 0x0000 | Turbidity Value | Float | 2 | R | 0 - range NTU |
| 0x0004 | Turbidity Factor | Float | 2 | RW | 0.1 to 9.99 |
| 0x0008 | Turbidity Offset | Float | 2 | RW | ±80% FS NTU |

### Configuration Registers (Function 0x03 read / 0x06 write)

| Address | Description | Type | Length | R/W | Notes |
|:--------|:------------|:-----|:-------|:----|:------|
| 0x000B | Wiper Time | Int | 1 | R | |
| 0x000C | Response Time | Int | 1 | RW | 3 to 60 seconds |
| 0x000D | Parameter | Int | 1 | R | 1 = turbidity, 2 = suspended solids |
| 0x000E | Probe Humidity | Int | 1 | R | Recommended < 10 |
| 0x0010 | Baud Rate | Int | 1 | RW | 0=4800, 1=9600, 2=19200, 3=38400 |
| 0x0011 | Slave Address | Int | 1 | RW | 1 to 200 |
| 0x0012 | Serial Number High | Int | 1 | R | First 4 digits |
| 0x0013 | Serial Number Low | Int | 1 | R | Last 4 digits |
| 0x0014 | Manual Wipe | Int | 1 | W | Send 66 to trigger |
| 0x0015 | Auto Wipe Interval | Int | 1 | W | Interval in minutes |

### Calibration Registers

| Address | Description | Type | Length | R/W | Notes |
|:--------|:------------|:-----|:-------|:----|:------|
| 0x003C | Factor 1 | Float | 2 | RW | Calibration factor 1 |
| 0x003E | Factor 2 | Float | 2 | RW | Calibration factor 2 |
| 0x0042 | Standard Solution Value | Float | 2 | RW | Write standard value for 2-point cal |

## 6. Reading Data

Read turbidity value (2 float registers from 0x0000):

```
Request:  01 03 00 00 00 02 C4 0B
Response: 01 03 04 XX XX XX XX [CRC]
```

**Decoding:**

| Register | Bytes | Format | Result |
|:---------|:------|:-------|:-------|
| 0x0000 to 0x0001 Turbidity | 4 bytes | IEEE-754 Float32 | **NTU** |

## 7. Address Configuration

### Change Address (0x01 → 0x02)

```
01 06 00 11 00 02 58 0E
```

### Change Baud Rate (9600 → 19200)

```
01 06 00 10 00 02 09 CE
```

### Broadcast Discovery

Use address 0xFE with only one device connected.

## 8. Calibration

The sensor is factory-calibrated. A two-point calibration using turbidity standard solutions is available for field recalibration.

### Two-Point Calibration Procedure

1. Connect sensor to PC via RS485-to-USB adapter.
2. Set parameters (select address 66) and clean the sensor lens.
3. Immerse sensor slowly in the **first** standard solution (recommended: 50 to 100 NTU range).
4. Write the standard value to register 0x0042. When the displayed turbidity value matches the input value, the first point is calibrated.
5. Clean and dry the sensor, then immerse in the **second** standard solution (recommended: 2000 to 2500 NTU range).
6. Write the standard value to register 0x0042. When the displayed turbidity value matches the input value, the second point is calibrated.

### Alternative: Direct Factor Adjustment

You can also directly write correction factors to registers 0x003C (Factor 1) and 0x003E (Factor 2) for fine-tuning the calibration curve.

### Calibration Notes

- Keep probe lens at least 15 cm from the bottom of the calibration cup
- Ensure no air bubbles on the lens surface
- Shield the calibration cup from ambient light during calibration

## 9. Installation Notes

**Placement**

- Three mounting options: quick-release poolside, classic poolside, or railing-mount (DN32 pipe)
- Submerge sensor fully in the measurement medium
- Avoid areas with excessive air bubbles or turbulence

**Maintenance**

- Clean the two optical lenses periodically
- Rinse with clean water, then wipe with detergent for stubborn deposits
- Inspect for damage; replace sealing rings annually
- Replace rubber wiper blade quarterly for sensors with self-cleaning
- Do NOT manually rotate the self-cleaning wiper  -  motor-driven only

## Ordering Information

| Part number | Description |
|:------------|:------------|
| AGRINOVO-TUR-110 | Self-cleaning Turbidity Sensor |

## Revision History

| Rev | Date | Changes |
|:----|:-----|:--------|
| 1.0 | June 2026 | Initial release |
| 1.1 | August 2026 | Layout revision |
| 1.2 | October 2026 | New layout |

## Contact

Agrinovo, https://agrinovo.io/products, contact@agrinovo.io, +972 54 688 8148 (direct / WhatsApp).